Difference between revisions of "MCE firmware"

From MCEWiki
(Hardware Peripherals)
(MCE Features and How-To Guides)
Line 43: Line 43:
 
** [[ Flux jumping | Flux Jumping]]
 
** [[ Flux jumping | Flux Jumping]]
 
** [[ FSFB Clamping Commands ]] (i.e. To Prevent Unlocked Pixels from Ramping, 5.x.x+)
 
** [[ FSFB Clamping Commands ]] (i.e. To Prevent Unlocked Pixels from Ramping, 5.x.x+)
 +
** [[ FSFB Servo Start Value ]]
  
 
* Address Card
 
* Address Card

Revision as of 17:17, 2 June 2010

Critical Notes

  • MCE firmware-load bug: In v5+ firmware, if you update firmware on any readout card, the card doesn't reply to any command unless the Clock card is also reprogrammed.
  • Always cleanly reset the MCE after a power-up or a reconfiguration! If not, MCE communication with the PC or Sync Box may not work.
> mce_reset_clean
  • After an mce_reset, not all of the DAC outputs are set to zero. In addition, values written to RAM-based registers are not reset. A full reconfiguration is required for a proper reset.
> rs psc cycle_pow 1
> mce_reset_clean

MCE Firmware Releases

Firmware Revisions:

Firmware Downloads and .JAM Conversions

MCE Features and How-To Guides

MCE Firmware Bugs/ Features Under Development

Hardware Peripherals

Software Tools

  • SVN links (We decided (sadly) NOT to migrate from CVS to SVN, if we ever do here is some information:)
SVN Manual
Accessing the SVN Repository
SVN Commit Guidelines
SVN Basic Operations
Tagging
Merging
Branching
Resurrecting Deleted Items

MCE Scripting