Difference between revisions of "Row-specific SQ2 feedback (fast switching)"

From MCEWiki
 
(4 intermediate revisions by the same user not shown)
Line 1: Line 1:
The SQ2-Feedback (SQ2FB) output of the MCE, controlled by a DAC can be set to either a constant value or a row-specific set of values switching at row-visit rate. The SQ2FB output is driven by the card plugged into the BC2 slot of the MCE. Either a Bias Card or an Address Card can be plugged into the BC2 slot.
+
The SQ2-Feedback (SQ2FB) output of the MCE, controlled by a DAC can be set to either a constant value or a row-specific set of values switching at row-visit rate. The SQ2FB output is driven by the card plugged into the BC2 slot of the MCE.
 
 
= Bias-card-driven SQ2FB =
 
  
 
== Hardware requirements ==
 
== Hardware requirements ==
Line 39: Line 37:
 
*** ''(In older mce_script, the equivalent is to set hardware_bac = 0)''
 
*** ''(In older mce_script, the equivalent is to set hardware_bac = 0)''
  
= Address-card-driven SQ2FB =
+
[[Category:Bias Card]]
 
+
[[Category:Bias Card Firmware]]
See older instructions here: [[ Fast SQ2 Feedback and TES Biasing with an Address Card ]]
 
 
 
== Hardware requirements ==
 
 
 
* No modifications are necessary to the address card.
 
* Note that the usual BC2 bias/heater line is not available when the biasing address card (BAC) is in use. However, if needed, the address-card hardware can be physically modified to provide the same functionality, for details see: http://www.phas.ubc.ca/~scuba2/sc2mce/system/addr_card/design/ECOs/AC-ECO049-20080508.doc
 
* The SQ2 FB DAC range is reduced from 16 to 14 bits when using the BAC.
 
 
 
== Firmware requirements ==
 
* AC 5.0.0 or later
 
;Relevant mce parameters
 
:{{param|ac|enbl_mux}}
 
:{{param|ac|const_val}}
 
:{{param|ac|fb_col0}}, {{param|ac|fb_col0|fb_col1}}, ... to {{param|ac|fb_col0|fb_col31}}
 
:{{param|ac|const_val39}}
 
:{{param|sys|row_len}}
 
 
 
== Software requirements ==
 
* mas/trunk : r439 or later
 
* mce_script/trunk : this has worked for a while...
 
 
 
;Configuration instructions
 
* When creating mce.cfg:
 
** $card_list must indicate a "bac" instead of a "bc2"
 
** mce_status -g should show entries for "bac fb_col#" and "sq2 fb_col#" and no entries for "bc2"
 
* In experiment.cfg:
 
** Set one of
 
*** config_fast_sq2 = 1;    # enable fast-SQ2 switching (and tuning)
 
*** config_fast_sq2 = 0;    # simulate non-fast-SQ2 bc2 by writing the same FB value for each row.
 
** Set
 
*** hardware_fast_sq2 = 1;  # biasing address card
 
*** ''(In older mce_script, set hardware_bac = 1 instead.)''
 
 
 
[[Category:Hardware]]
 
[[Category:Firmware]]
 

Latest revision as of 19:04, 2 August 2018

The SQ2-Feedback (SQ2FB) output of the MCE, controlled by a DAC can be set to either a constant value or a row-specific set of values switching at row-visit rate. The SQ2FB output is driven by the card plugged into the BC2 slot of the MCE.

Hardware requirements

  • Bias cards prior to Rev. F have a slow opamp at the output followed by 2 sets of kHz-range RC filters that limit the output switching frequency to few kHz range and prevents the output from switching at the row-visit rate, typically around 500kHz. Therefore, when these bias cards are used in BC2 slot, only a constant SQ2FB value (one value per array-column) can be specified.
  • Bias Cards Rev. F and later have faster outputs and they can be switched at row-visit rates up to about 900kHz .

Firmware requirements

The firmware support for switching SQ2FB values at row-visit rate is added starting revision BC 5.0.5.

Relevant mce parameters
  • enbl_mux : one value per channel, to enable / disable multiplexing DAC values. Default: mode = 0 (disabled); write 1 to enable.
    • When enbl_mux = 0, DACs are loaded with values specified using flux_fb parameter.
    • When enbl_mux = 1, multiplexing is enabled and DACs are loaded with a new value from fb_col# at each row switch.
  • flux_fb: values for each of the 32 flux-feedback 16-bit DACs on Bias Card. These are used for any column having enbl_mux = 0.
  • fb_col0, fb_col1, ... to fb_col31: registers, each of which may contain up to 41 sq2fb values. These are only used when enbl_mux = 1 for the column

Muxing can be enabled / disabled on each column independently. For classic, constant FB output, ensure enbl_mux = 0 0 0 0 ... .

Software requirements

  • mas/trunk : r510 or later - servo programs will automatically detect the fast bias card.
  • mce_script/trunk: r764 or later
Configuration instructions
  • When creating mce.cin:
    • $fw_rev["bc"] = 0x5000005 or greater.
    • $card_list should indicate a "bc2" (i.e. not a "bac")
    • mce_status -g should show entries for "bc2 fb_col#" and "sq2 fb_col#" as well as "bc2 flux_fb" and "sq2 fb".
  • In experiment.cfg:
    • Set one of
      • config_fast_sq2 = 1; # enable fast-SQ2 switching (and tuning)
      • config_fast_sq2 = 0; # disable fast-SQ2 switching
    • Set
      • hardware_fast_sq2 = 2; # new bias card, supports fast- and no-switching
      • (In older mce_script, the equivalent is to set hardware_bac = 0)