Difference between revisions of "Sync Box"

From MCEWiki
(Schematics)
Line 11: Line 11:
 
:Each Fibre transmitter (HFBR1119) consumes 185mA x 5V = 0.925W. To lower consumption, the unused transmitters can be removed.
 
:Each Fibre transmitter (HFBR1119) consumes 185mA x 5V = 0.925W. To lower consumption, the unused transmitters can be removed.
  
;Dimensions
+
;Dimensions (approximate)
 +
: 9cm x 21cm x 48cm for rack-mount AC-in
 +
: 5cm x 20cm x 25cm for small DC-in
 
;Weight
 
;Weight
 
: A rack-mount AC-in Sync box unit weighs 1.9kg.
 
: A rack-mount AC-in Sync box unit weighs 1.9kg.
 
;Inputs/Outputs
 
;Inputs/Outputs
: [[Sync_Box_AC-in_Rack_Mount_io |rack-mount AC-in Sync Box]] [[Image:Sync_Box_AC-In_Rack_Mount.JPG|20px]]  
+
: [[Sync_Box_AC-in_Rack_Mount_io |IO for rack-mount AC-in Sync Box]] [[Image:Sync_Box_AC-In_Rack_Mount.JPG|20px]]  
: [[Sync_Box_DC-in io| small DC-in Sync Box]]  
+
: [[Sync_Box_DC-in io| IO for small DC-in Sync Box]]  
 
: [[Pin Translation table between Rack-mount and DC-in versions of Sync Box]]
 
: [[Pin Translation table between Rack-mount and DC-in versions of Sync Box]]
 
== User's Guide ==
 
== User's Guide ==

Revision as of 12:23, 15 May 2013

Template:Hierarchy header The Sync Box generates a data-valid pulse along with a sequential 32-bit number to synchronize the data-acquisition operation of up-to 8 MCE subracks and other housekeeping equipment. The 32-bit number can be used to stamp all collected data points. The Sync Box has 8 fibre outputs to connect to the MCE subracks and few TTL/RS485 outputs to be used by other housekeeping equipment.

The Sync Box comes in 2 different packaging or enclosures:

  1. A rack-mount AC-in Sync Box Sync Box AC-In Rack Mount.JPG Sync Box AC-In Rack Mount Front.JPG
  2. A DC-in 20cmx20cmx5cm Sync Box Sync Box DC-In Chassis.JPG

Specification

Power consumption
A Sync box equipped with all 8 Fibre transmitters consumes 8.75W.
Each Fibre transmitter (HFBR1119) consumes 185mA x 5V = 0.925W. To lower consumption, the unused transmitters can be removed.
Dimensions (approximate)
9cm x 21cm x 48cm for rack-mount AC-in
5cm x 20cm x 25cm for small DC-in
Weight
A rack-mount AC-in Sync box unit weighs 1.9kg.
Inputs/Outputs
IO for rack-mount AC-in Sync Box Sync Box AC-In Rack Mount.JPG
IO for small DC-in Sync Box
Pin Translation table between Rack-mount and DC-in versions of Sync Box

User's Guide

The functional description of Sync Box is available here.

Note that the user's guide describes the original Rack-mount AC-in enclosure and the document is not yet up to date for the DC-in enclosure.


Schematics

Firmware

For instructions and downloads see: Sync Box Firmware

Setting up the Sync Box

You can use an rs232 terminal emulator (e.g., minicom on a linux PC, see below) to issue commands to the Sync Box, a list of all the commands available is given by typing

> h

Setting up the MCE to use the Sync Box

To check that the sync box is connected correctly to the MCE, issue the command

wb cc select_clk 1

and then read back the value:

rb cc select_clk
Line   0 : ok : 1

Setting select_clk to 1 tells the MCE to sync its internal clock with the clock signal encoded from the sync box. If the MCE does not see the sync box, the parameter will revert to 0 in about 1 second.

To accept sync triggers and serial numbers from the sync box, two more commands are needed:

wb cc use_sync 2
wb cc use_dv 2

These registers are configured automatically by the config system if the parameters "hardware_sync" and "config_sync" are both 1 in experiment.cfg. If you change the values in $MAS_DATA/experiment.cfg instead of $MAS_TEMPLATE/experiment.cfg, remember to "mce_make_config" before re-running the config script.

Also be sure that you have copied the correct mce.cfg file over to /etc/mce folder from the following locations:

  • /home/mce/mas_trunk/config/v4 (for 4-series firmware), or
  • /home/mce/mas_trunk/config/v5 (for 5-series firmware)

A quick way to check that sync numbers are getting into the MCE frame data is to take some frames and check the data using the eat_packets tool:

mce@mce-ubc-2:~$ mce_status -s | grep fw_rev

mce@mce-act-b1$  mce_run sync_test 400 s
RUNFILE_NAME=/data/cryo/current_data//sync_test.run
FRAME_BASENAME=/data/cryo/current_data//sync_test

mce@mce-ubc-2:~/mce_script/script$ eat_packets -n 5424 -f $MAS_DATA/mce_run10
Forced     frame_size=1356 (5424)
offset     frm_idx   frame#
0000000000        0        0  surprise sync_dv 4168076, after sequence [0,0)
EOF, exiting after 10000 frames + 0 bytes

If the mce_run does not terminate quickly, the sync box may not be properly connected to the MCE. If the eat_packets complains only about the first frame (offset 00000000), then the sync numbers are intact and incrementing contiguously. If eat_packets complains about pretty much every frame, the sync numbers are not getting into the frame data.

minicom Setup

To issue commands to a Sync Box over an RS-232 line, you can use a Linux program called 'MiniCom'. To install this application, type the following:

> sudo apt-get install minicom

To start MiniCom:

> sudo minicom

Minicom needs the following options changed for it to work properly:

> <ctrl-a>, <z>, <a>
> <ctrl-a>, <z>, <o>, <Serial port setup> <Serial Device> = /dev/ttyS0
> <ctrl-a>, <z>, <o>, <Serial port setup> <Bps/Par/Bits> = 9600 8N1
> <ctrl-a>, <z>, <o>, <Serial port setup> <Hardware Flow Control> = No
> <ctrl-a>, <z>, <o>, <Serial port setup> <Software Flow Control> = No
> <ctrl-a>, <z>, <o>, <Modem and dialing> <Modem has DCD line> = No

After you modify these settings, remember to save, exit, and restart Minicom.